Test ID: MEX
Mexiletine, Serum
Secondary ID
A test code used for billing and in test definitions created prior to November 2011
NY State Approved
Indicates the status of NY State approval and if the test is orderable for NY State clients.
Useful For
Suggests clinical disorders or settings where the test may be helpful
Assessing achievement of optimal therapeutic concentrations
Assessing potential toxicity
Method Name
A short description of the method used to perform the test
High-Performance Liquid Chromatography (HPLC)
Reporting Name
A shorter/abbreviated version of the Published Name for a test; an abbreviated test name
Aliases
Lists additional common names for a test, as an aid in searching
Specimen Type
Describes the specimen type needed for testing
Specimen Required
Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.
Container/Tube: Red top
Specimen Volume: 3 mL
Collection Instructions: Draw blood after patient has been receiving mexiletine for at least 3 days and just before administration of the next dose.
Specimen Minimum Volume
Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.
Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected
| Hemolysis | Mild OK; Gross reject |
| Lipemia | Mild OK; Gross reject |
| Icterus | Mild OK; Gross reject |
| Other | Plasma, serum gel tube |
Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.
| Specimen Type | Temperature | Time |
|---|---|---|
| Serum Red | Refrigerated (preferred) | 14 days |
| Frozen | 14 days | |
| Ambient | 7 days |
Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test
Mexiletine is a class I B antiarrhythmic with electrophysiologic properties similar to lidocaine and is useful in suppression of ventricular arrhythmias.
The drug exhibits a high degree of oral bioavailability, is approximately 60% protein bound, and undergoes renal clearance at a rate of 10.3 mL/min/kg. Mexiletine has a volume of distribution of 9.5 L/kg at a half-life of 11 hours. Myocardial infarction and uremia reduce the rate of clearance and increase the half-life of mexiletine, requiring dosage adjustment guided by drug monitoring.
Mexiletine toxicity occurs at concentrations >2.0 mcg/mL (trough value) and is characterized by symptoms of nausea, hypotension, sinus bradycardia, paresthesia, seizures, intermittent left bundle branch block, and temporary asystole.
Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.
Therapeutic concentration: 0.8-2.0 mcg/mL (trough value)
Toxic concentration: >2.0 mcg/mL (trough value)
Interpretation
Provides information to assist in interpretation of the test results
Optimal response to mexiletine occurs when the serum concentration is within the range of 0.8 to 2.0 mcg/mL (trough value).
Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances
No significant cautionary statements
Clinical Reference
Provides recommendations for further in-depth reading of a clinical nature
Joseph SP, Holt DW: Electrophysiological properties of mexiletine assessed with respect to plasma concentrations. Eur J Cardiol 1980;11:115-121
Method Description
Describes how the test is performed and provides a method-specific reference
Mexiletine is extracted with an organic solvent from serum at neutral or basic pH, but pH 10.4 is optimal for minimal interference from other drugs and maximal recovery. The organic extract is acidified and then evaporated under a stream of N(2). The residue is taken up in mobile phase and chromatographed by reverse-phase high-performance liquid chromatography (HPLC). Mexiletine is detected by absorbance at 210 nm and quantified by peak height ratios relative to an internal standard, p-chlorodisopyramide.(Mastropaolo W, Holmes DR, Osborn MJ, et al: Improved liquid-chromatographic determination of mexiletine, an antiarrhythmic drug, in plasma. Clin Chem 1984;30:319-322)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Monday, Wednesday, Friday; 9:00 a.m.
Analytic Time
Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.
Maximum Laboratory Time
Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result
Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded
Performing Laboratory Location
The location of the laboratory that performs the test
Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.
CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.
80299
LOINC® Code Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.
| Result ID | Reporting Name | LOINC Code |
|---|---|---|
| 9245 | Mexiletine, S | 3819-0 |