Test ID: CDOMB
Cadmium for Occupational Monitoring, Blood
Secondary ID
A test code used for billing and in test definitions created prior to November 2011
NY State Approved
Indicates the status of NY State approval and if the test is orderable for NY State clients.
Useful For
Suggests clinical disorders or settings where the test may be helpful
Detecting exposure to cadmium
Special Instructions and Forms
Describes specimen collection and preparation information, test algorithms, and other information pertinent to test. Also includes pertinent information and consent forms to be used when requesting a particular test
Method Name
A short description of the method used to perform the test
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
Reporting Name
A shorter/abbreviated version of the Published Name for a test; an abbreviated test name
Aliases
Lists additional common names for a test, as an aid in searching
Cadmium (Cd)
Specimen Type
Describes the specimen type needed for testing
Specimen Required
Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.
Container/Tube: Royal, blue-top Monoject trace element blood collection tube, product #8881-307022 (Supply T183), containing EDTA as an anticoagulant
Specimen Volume: Full tube
Collection Instructions:
1. Send specimen in original tube.
2. See Metals Analysis-Collection and Transport in Special Instructions for complete instructions.
Additional Information: High concentrations of gadolinium and iodine are known to interfere with most metals tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Specimen Minimum Volume
Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.
Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected
| Hemolysis | Mild OK; Gross OK |
| Lipemia | Mild OK; Gross OK |
| Icterus | Mild OK; Gross reject |
| Other | Green-top (heparin) tube |
Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.
| Specimen Type | Temperature | Time |
|---|---|---|
| Whole blood | Refrigerated (preferred) | 21 days |
| Ambient | 21 days | |
| Frozen | 21 days |
Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test
The toxicity of cadmium resembles the other heavy metals (arsenic, mercury, and lead) in that it attacks the kidney; renal dysfunction with proteinuria with slow onset (over a period of years) is the typical presentation. Breathing the fumes of cadmium vapors leads to nasal epithelial deterioration and pulmonary congestion resembling chronic emphysema.
The most common source of chronic exposure comes from spray painting of organic-based paints without use of a protective breathing apparatus; auto repair mechanics represent a susceptible group for cadmium toxicity. Tobacco smoke is another common source of cadmium exposure.
Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.
0.0-4.9 mcg/L
Reference values apply to all ages.
Interpretation
Provides information to assist in interpretation of the test results
Normal blood cadmium is <5.0 mcg/L, with most results in the range of 0.5 mcg/L to 2.0 mcg/L.
Acute toxicity will be observed when the blood level exceeds 50 mcg/L.
Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances
High concentrations of gadolinium and iodine are known to interfere with most metals tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen cannot be collected for 96 hours.
Clinical Reference
Provides recommendations for further in-depth reading of a clinical nature
1. Moreau T, Lellouch J, Juguet B, et al: Blood cadmium levels in a general male population with special reference to smoking. Arch Environ Health 1983;38:163-167
2. Occupational Safety and Health Administration, US Department of Labor: Cadmium Exposure Evaluation. Updated 9/2/2008. Available from URL:osha.gov/SLTC/cadmium/evaluation.html
Method Description
Describes how the test is performed and provides a method-specific reference
Cadmium is analyzed on an inductively coupled plasma-mass spectrometer. Aqueous acidic calibrating standards and blanks are diluted with an aqueous acidic diluent containing 3 internal standards. Quality control specimens and patient urines are also diluted in an identical manner. In turn, all diluted blanks, calibrating standards, quality control specimens, and patient specimens are aspirated into a pneumatic nebulizer and the resulting aerosol directed to the hot plasma discharge by a flow of argon. In the annular plasma the aerosol is vaporized, atomized, and then ionized. The ionized gases plus neutral species formed in the annular plasma space are aspirated from the plasma through an orifice into a quadrapole mass spectrometer. The mass range from 1 amu to 263 amu is rapidly scanned multiple times and ion counts tabulated for each mass of interest. Instrument response is defined by the linear relationship of analyte concentration versus ion count ratio (analyte ion count/internal standard ion count). Analyte concentrations are derived by reading the ion count ratio for each mass of interest and determining the concentration from the response line.(Nixon DE, Moyer TP: Routine clinical determination of lead, arsenic, cadmium, and thallium in urine and whole blood by inductively coupled plasma-mass spectrometry. Spectrochimica Acta Part B-Atomic Spectroscopy 1996;51:13-25)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Monday through Friday; 8 a.m. - 2 p.m.; Saturday; 8 a.m. - 3 p.m. Continuously
Analytic Time
Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.
Maximum Laboratory Time
Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result
Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded
Performing Laboratory Location
The location of the laboratory that performs the test
Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.
CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.
82300
LOINC® Code Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.
| Result ID | Reporting Name | LOINC Code |
|---|---|---|
| 89539 | Cadmium Occupational Monitor, B | 5609-3 |