Mobile Site ›
Normal View

Test ID: CAFF
Caffeine, Serum

Secondary ID A test code used for billing and in test definitions created prior to November 2011

8754

NY State Approved Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Useful For Suggests clinical disorders or settings where the test may be helpful

Monitoring therapy in neonates

 

Assessing toxicity in neonates

Method Name A short description of the method used to perform the test

Immunoassay

Reporting Name A shorter/abbreviated version of the Published Name for a test; an abbreviated test name

Caffeine, S

Aliases Lists additional common names for a test, as an aid in searching

Caffeine

Specimen Type Describes the specimen type needed for testing

Serum Red

Specimen Required Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.

Container/Tube: Red top

Specimen Volume: 1 mL

Specimen Minimum Volume Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.

0.4 mL

Reject Due To Identifies specimen types and conditions that may cause the specimen to be rejected

Hemolysis

Mild OK; Gross reject

Lipemia

Mild OK; Gross reject

Icterus

Mild OK; Gross reject

Other

Serum gel tube

Specimen Stability Information Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.

Specimen TypeTemperatureTime
Serum RedRefrigerated (preferred)14 days
 Ambient 14 days
 Frozen 14 days

Clinical Information Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Caffeine is used to treat apnea that occurs in newborn infants, the most frequent complication seen in the neonatal nursery.

  

Caffeine is administered orally (nasogastric tube) as a loading dose of 3 mg/kg followed by a maintenance dose of 1 mg/kg administered once every 24 to 48 hours, depending on the patient's response and the serum level.

 

In neonates, caffeine has a half-life that ranges from 20 to 100 hours, which is much longer than in adults (typically 4-6 hours) due to the immaturity of the neonatal liver. This requires that small doses be administered at much longer intervals than would be predicted based on adult pharmacokinetics.

 

The volume of distribution of caffeine is 0.6 L/kg and the drug is approximately 35% protein bound.

 

Toxicity observed in neonates is characterized by central nervous system and skeletal muscle stimulation and bradycardia. These symptoms are seen in adults at lower levels than in neonates, suggesting that neonates have much greater tolerance to the drug.

Reference Values Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Therapeutic concentration: 5-15 mcg/mL

Toxic concentration: > or =30 mcg/mL

Interpretation Provides information to assist in interpretation of the test results

Optimal pharmacologic response occurs when the serum level is in the range of 5 mcg/mL to 15 mcg/mL.

 

Toxicity in neonates and adults may be seen when the serum level is > or =20 mcg/mL.

Cautions Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

This assay is not intended to detect levels in adults.

Clinical Reference Provides recommendations for further in-depth reading of a clinical nature

Ou CN, Frawley VL: Concurrent measurement of theophylline  and caffeine  in  neonates by an interference-free liquid-chromatographic method. Clin Chem 1983;29:1934-1936

Method Description Describes how the test is performed and provides a method-specific reference

Enzyme-multiplied immunoassay technique (EMIT) performed using Olympus analyzer:EMIT offers an alternative to the traditional spectroscopic and chromatographic method for quantitating blood concentrations of drugs. The technique for drugs is based upon an enzymatic assay for glucose-6-phosphate dehydrogenase, using spectral properties at 340 nm, in which the reduction of nicotinamide adenine dinucleotide (NAD) substrate is monitored. The basis of the

drug detection technique is an immunological reaction between the drug and a specific antibody. The reagent contains the enzyme (glucose-6-phosphate dehydrogenase) to which the drug is covalently bound and antibody-specific to the drug. The antibody binds most of the drug-bound enzyme, rendering the enzyme inactive. This results in a baseline enzymatic activity. In the presence of free drug, antibody equilibrates between free drug and enzyme-bound drug leaving some of the drug-bound enzyme uncomplexed and able to catalyze the reaction. If more free drug is introduced, either as standard or sample, then competition for the antibody takes place between the drug in the sample and the drug attached to the enzyme. This results in more drug-bound enzyme being left uncomplexed and able to catalyze the enzyme reaction at a greater rate as compared to the baseline activity. The observed enzyme activity increases with the amount of total free drug in the sample. (Moyer TP: Therapeutic Drug Monitoring. In Tietz Textbook of Clinical Chemistry. 4th edition. Edited by CA Burtis, ER Ashwood. Philadelphia, WB Saunders Company, 2005, pp 1237-1285)

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday through Sunday; Varies

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

Same day/1 day

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result

1 day

Specimen Retention Time Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

14 days

Performing Laboratory Location The location of the laboratory that performs the test

Rochester

Test Classification Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.

This test has been cleared or approved by the U.S. Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

80299

LOINC® Code Information Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.

Result IDReporting NameLOINC Code
8754Caffeine, S3422-3