Test ID: DIG
Digoxin, Serum
Secondary ID
A test code used for billing and in test definitions created prior to November 2011
NY State Approved
Indicates the status of NY State approval and if the test is orderable for NY State clients.
Useful For
Suggests clinical disorders or settings where the test may be helpful
Monitoring digoxin therapy
Method Name
A short description of the method used to perform the test
Electrochemiluminescent Immunoassay
Reporting Name
A shorter/abbreviated version of the Published Name for a test; an abbreviated test name
Aliases
Lists additional common names for a test, as an aid in searching
Specimen Type
Describes the specimen type needed for testing
Specimen Required
Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.
Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Specimen Volume: 1 mL
Specimen Minimum Volume
Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.
Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected
| Hemolysis | Mild OK; Gross reject |
| Lipemia | NA |
| Icterus | NA |
| Other | NA |
Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.
| Specimen Type | Temperature | Time |
|---|---|---|
| Serum | Refrigerated (preferred) | 7 days |
| Frozen | 180 days |
Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test
Compounds in the digitalis family of glycosides consist of a steroid nucleus, a lactone ring, and a sugar. Digoxin is widely prescribed for the treatment of congestive heart failure and various disturbances of cardiac rhythm. Digoxin improves the strength of myocardial contraction and results in the beneficial effects of increased cardiac output, decreased heart size, decreased venous pressure, and decreased blood volume. Digoxin therapy also results in stabilized and slowed ventricular pulse rate. These therapeutic effects are produced through a network of direct and indirect interactions upon the myocardium, blood vessels, and the autonomic nervous system.
Digoxin is well absorbed after oral administration and is widely distributed to tissues, especially the heart, kidney, and liver. A number of factors can alter normal absorption, distribution, and bioavailability of the drug, including naturally occurring enteric bacteria in the bowel, presence of food in the gut, strenuous physical activity, ingestion of quinine or quinidine, and concomitant use of a wide range of drugs. Children generally require higher concentrations of digoxin.
After oral administration, there is an early rise in serum concentration. Equilibration of serum and tissue levels occurs at approximately 6 to 8 hours. For this reason, blood specimens for digoxin analysis should be drawn at least 6 to 8 hours after drug administration. Digoxin is excreted primarily in the urine. The average elimination half-life is 36to 40 hours, but may be considerably prolonged in those with renal disease, causing digoxin accumulation and toxicity.
Symptoms of digoxin toxicity often mimic the cardiac arrhythmia's for which the drug was originally prescribed (eg, heart block and heart failure). Other typical symptoms of toxicity include gastrointestinal effects, including anorexia, nausea, vomiting, abdominal pain and diarrhea, and neuropsychologic symptoms, such as fatigue, malaise, dizziness, clouded or blurred vision, visual and auditory hallucination, paranoid ideation, and depression. Toxicity of digoxin may reflect several factors: the drug has a narrow therapeutic window (a very small difference exists between therapeutic and toxic tissue levels); individuals vary in their ability to metabolize and respond to digoxin; absorption of various oral forms of digoxin may vary over a 2-fold range; susceptibility to digitalis toxicity apparently increases with age.
Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.
Therapeutic concentration: 0.5-2.0 ng/mL
Toxic concentration: > or =4.0 ng/mL
Pediatric toxic concentrations may be higher.
Reference values have not been established for patients that are <16 years of age.
Interpretation
Provides information to assist in interpretation of the test results
The therapeutic range is 0.5 to 2.0 ng/mL.
Levels >4.0 ng/mL may be potentially life-threatening.
Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances
Patients vary in their responsiveness to the drug, and renal dysfunction permits accumulation of digoxin in the serum.
This assay measures both bound and free digoxin, so high values will be found in patients undergoing treatment with digibind (antibody fragment therapy).
"Digoxin-like" immunoreactive factors may cause falsely-elevated values in some neonates and patients with advanced liver or renal disease.(3)
In patients undergoing therapy with high biotin doses (>5 mg/day), do not draw specimens until at least 8 hours have elapsed since the last biotin administration.
Clinical Reference
Provides recommendations for further in-depth reading of a clinical nature
1. Applied Therapeutic Drug Monitoring. Vol 2. Edited by TP Moyer, RL Boeckx. Washington, DC, American Association for Clinical Chemistry, 1984
2. Jortani SA, Voldew R Jr: Digoxin and its related endogenous factors. Crit Rev Clin Lab Sci 1997;34:225-274
3. Datta P, Hinz V, Klee G: Comparison four digoxin immunoassays with respect to interference from digoxin-like immunoreactive factors. Clin Biochem 1996;29(6):541-547
Method Description
Describes how the test is performed and provides a method-specific reference
The E Modular digoxin method is a competitive electrochemiluminescence immunoassay, which employs a monoclonal antibody, directed against digoxin. Digoxin in the specimen competes with the added digoxin derivative labeled with biotin for the binding sites on the ruthenylated antibody-complex. Streptavidin-coated microparticles are added and the mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Application of voltage to the electrode induces the chemiluminescent emission, which is then measured. (Package insert: E Modular, DIGOXIN. Roche Diagnostics Corporation, Indianapolis IN)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Monday through Sunday; Continuously
Analytic Time
Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.
Maximum Laboratory Time
Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result
Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded
Performing Laboratory Location
The location of the laboratory that performs the test
Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.
CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.
80162
LOINC® Code Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.
| Result ID | Reporting Name | LOINC Code |
|---|---|---|
| DIG | Digoxin, S | 10535-3 |


