Test ID: CUS
Copper, Serum
Secondary ID
A test code used for billing and in test definitions created prior to November 2011
NY State Approved
Indicates the status of NY State approval and if the test is orderable for NY State clients.
Useful For
Suggests clinical disorders or settings where the test may be helpful
Diagnosis of:
-Wilson disease
-Primary biliary cirrhosis
-Primary sclerosing cholangitis
Special Instructions and Forms
Describes specimen collection and preparation information, test algorithms, and other information pertinent to test. Also includes pertinent information and consent forms to be used when requesting a particular test
Method Name
A short description of the method used to perform the test
Dynamic Reaction Cell-Inductively Coupled Plasma-Mass Spectrometry (DRC-ICP-MS)
Reporting Name
A shorter/abbreviated version of the Published Name for a test; an abbreviated test name
Aliases
Lists additional common names for a test, as an aid in searching
Cu (Copper)
Kaiser Fleischer Ring
Wilson's Disease
Specimen Type
Describes the specimen type needed for testing
Specimen Required
Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.
Collection Container/Tube: Plain, royal-blue top Monoject trace element blood collection tube, product #8881-307006 (Supply T184)
Submission Container/Tube: 7-mL Mayo metal-free, screw-capped, polypropylene vial (Supply T173)
Specimen Volume: 0.8 mL
Collection Instructions:
1. Allow the specimen to clot for 30 minutes; then centrifuge the specimen to separate serum from the cellular fraction.
2. Remove the stopper. Carefully pour specimen into Mayo metal-free, polypropylene vial, avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer.
3. See Metals Analysis-Collection and Transport in Special Instructions for complete instructions.
Additional Information: High concentrations of gadolinium, iodine, and barium are known to interfere with most metals tests. If either gadolinium-, iodine-, or barium-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Specimen Minimum Volume
Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.
Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected
| Hemolysis | Mild OK; Gross OK |
| Lipemia | Mild OK; Gross OK |
| Icterus | Mild OK; Gross reject |
| Other | NA |
Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.
| Specimen Type | Temperature | Time |
|---|---|---|
| Serum | Refrigerated (preferred) | 14 days |
| Ambient | 14 days | |
| Frozen | 14 days |
Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test
In serum from normal, healthy humans, more than 95% of the copper is incorporated into ceruloplasmin; the remaining copper is loosely bound to albumin.
Low serum copper, most often due to excess iron or zinc ingestion and infrequently due to dietary copper deficit, results in severe derangement in growth and impaired erythropoiesis. Low serum copper is also observed in hepatolenticular degeneration (Wilson disease) due to a decrease in the synthesis of ceruloplasmin and allelic variances in cellular metal ion transporters. In Wilson disease, the albumin-bound copper may actually be increased, but ceruloplasmin copper is low, resulting in low serum copper. However, during the acute phase of Wilson disease (fulminant hepatic failure), ceruloplasmin and copper may be normal; in this circumstance, hepatic inflammation causes increased release of ceruloplasmin. It is useful to relate the degree of liver inflammation to the ceruloplasmin and copper-see discussion on hypercupremia below. Significant hepatic inflammation with normal ceruloplasmin and copper suggest acute Wilson disease.
Other disorders associated with decreased serum copper concentrations include malnutrition, hypoproteinemia, malabsorption, nephrotic syndrome, Menkes disease, copper toxicity, and megadosing of zinc-containing vitamins (zinc interferes with normal copper absorption from the gastrointestinal tract).
Hypercupremia is found in primary biliary cirrhosis, primary sclerosing cholangitis, hemochromatosis, malignant diseases (including leukemia), thyrotoxicosis, and various infections. Serum copper concentrations are also elevated in patients taking contraceptives or estrogens and during pregnancy.
Since the gastrointestinal (GI) tract effectively excludes excess copper, it is the GI tract that is most affected by copper ingestion. Increased serum concentration does not, by itself, indicate copper toxicity.
Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.
0-2 months: 0.40-1.40 mcg/mL
3-6 months: 0.40-1.60 mcg/mL
7-9 months: 0.40-1.70 mcg/mL
10-12 months: 0.80-1.70 mcg/mL
13 months-10 years: 0.80-1.80 mcg/mL
> or =11 years: 0.75-1.45 mcg/mL
Interpretation
Provides information to assist in interpretation of the test results
Serum copper below the normal range is associated with Wilson disease, as well as a variety of other clinical situations (see Clinical Information). Excess use of denture cream containing zinc can cause hypocupremia.
Serum concentrations above the normal range are seen in primary biliary cirrhosis and primary sclerosing cholangitis, as well as a variety of other clinical situations (see Clinical Information).
Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances
High concentrations of gadolinium, iodine, and barium are known to interfere with most metals tests. If either gadolinium, iodine, or barium-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Clinical Reference
Provides recommendations for further in-depth reading of a clinical nature
1. McCullough AJ, Fleming CR, Thistle JL, et al: Diagnosis of Wilson's disease presenting as fulminant hepatic failure. Gastroenterology 1983;84:161-167
2. Wiesner RH, LaRusso NF, Ludwig J, Dickson ER: Comparison of the clinicopathologic features of primary sclerosing cholangitis and primary biliary cirrhosis. Gastroenterology 1985;88:108-114
3. Spain RI, Leist TP, De Sousa EA. When metals compete: a case of copper-deficiency myeloneuropathy and anemia. Nat Clin Pract Neurol 2009 Feb;5(2):106-111
4. Kale, SG, Holmes CS, Goldstein DS, et al: Neonatal Diagnosis and Treatment of Menkes Disease. N Engl J Med 2008 Feb 7;358(6):605-614
5. Nations SP, Boyer PJ, Love LA, et al: Denture cream: An unusual source of excess zinc, leading to hypocupremia and neurologic disease. Neurology 2008;71;639-643
Method Description
Describes how the test is performed and provides a method-specific reference
This assay is performed on an inductively coupled plasma-mass spectrometer in dynamic reaction cell mode. Calibrating standards and blanks are diluted with an aqueous acidic diluent containing internal standard(s). Quality control specimens and patient samples are diluted in an identical manner. In turn, all diluted blanks, calibrating standards, quality control specimens and patient specimens are aspirated into a pneumatic nebulizer and the resulting aerosol directed to the hot plasma discharge by a flow of argon. In the annular plasma the aerosol is vaporized, atomized, then ionized. The ionized gases plus neutral species formed in the annular plasma space are aspirated from the plasma through an orifice into a quadrupole mass spectrometer. The mass range from 1 to 263 amu is rapidly scanned multiple times and ion counts tabulated for each mass of interest. Instrument response is defined by the linear relationship of analyte concentration vs. ion count ratio (analyte ion count/internal standard ion count). Analyte concentrations are derived by reading the ion count ratio for each mass of interest and determining the concentration from the response line. (Hanley MM, Murray DL, Moyer TP, et al: Selenium, Copper and Zinc Serum: Attenuation of Interferences by Collision/Reaction Cell ICP-MS [Abstract WP22]. 2012 Winter Conference on Plasma Spectrochemistry, Tucson, Arizona, January 9-14, 2012)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Monday through Friday; 8 a.m. - 6 p.m.; Saturday; 8 a.m. - 3 p.m./Continuously
Analytic Time
Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.
Maximum Laboratory Time
Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result
Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded
Performing Laboratory Location
The location of the laboratory that performs the test
Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.
CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.
82525
LOINC® Code Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.
| Result ID | Reporting Name | LOINC Code |
|---|---|---|
| 8612 | Copper, S | 5631-7 |


