Mobile Site ›
Normal View

Test ID: PNRP
Pneumocystis jiroveci, Molecular Detection, PCR

Secondary ID A test code used for billing and in test definitions created prior to November 2011

81698

NY State Approved Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Useful For Suggests clinical disorders or settings where the test may be helpful

Preferred test for detection of Pneumocystis

Method Name A short description of the method used to perform the test

Real-Time Polymerase Chain Reaction (PCR)
(PCR is utilized pursuant to a license agreement with Roche Molecular Systems, Inc)

Reporting Name A shorter/abbreviated version of the Published Name for a test; an abbreviated test name

Pneumocystis PCR

Specimen Type Describes the specimen type needed for testing

Varies

Specimen Required Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.

The high sensitivity of amplification by PCR requires the specimen to be processed in an environment in which contamination of the specimen by Pneumocystis species DNA is not likely.

 

Forms: If not ordering electronically, submit a Microbiology Request Form (Supply T244) with the specimen.

 

Acceptable Specimens: Cerebrospinal fluid, respiratory specimens (eg, bronchoalveolar lavage fluid, bronchial washing, sputum), pleural fluid, respiratory tissue (fresh)

 

Unacceptable Specimens: Body fluid other than cerebrospinal fluid or pleural fluid, blood, bone, non-respiratory tissue, bone marrow, organ tissues other than lung, paraffin-embedded tissue, specimen in anaerobe vial or viral transport medium (including but not limited to M4, M5, BD viral transport media, thioglycolate broth), stool, swab, tissue in formalin fluid, urine, specimen >7 days old

 

Specimen must arrive within 7 days of collection; specimen >7 days will be rejected.

 

Specimen source is required.

 

Submit only 1 of the following specimens:

 

Specimen Type: Body fluid

Sources: Cerebrospinal or pleural

Container/Tube: Sterile container

Specimen Volume: 1 mL

                                                          

Specimen Type: Respiratory

Sources: Bronchoalveolar lavage fluid, bronchial washing, or sputum

Container/Tube: Sterile container

Specimen Volume: 1 mL

 

Specimen Type: Tissue

Sources: Respiratory

Container/Tube: Sterile container

Specimen Volume: 5-10 mm

Collection Instructions: Submit fresh tissue.

Specimen Minimum Volume Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.

0.5 mL

Reject Due To Identifies specimen types and conditions that may cause the specimen to be rejected

Hemolysis

NA

Lipemia

NA

Icterus

NA

Other

Body fluid other than cerebrospinal fluid or pleural fluid, blood, bone, non-respiratory tissue, bone marrow, organ tissues other than lung, paraffin-embedded tissue, specimen in anaerobe vial or viral transport medium, stool, swab, tissue in formalin fluid, urine, specimen >7 days old

Specimen Stability Information Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.

Specimen TypeTemperatureTime
VariesRefrigerated (preferred)7 days
 Frozen 7 days

Clinical Information Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Pneumocystis pneumonia is an important cause of opportunistic infection in immunocompromised patients, particularly those with HIV. The causative agent, Pneumocystis jiroveci, cannot be cultured in vitro and, therefore, laboratory detection has historically relied upon microscopic identification directly from patient specimens using fluorescent stains or antibodies. Unfortunately, stains often lack sensitivity and require expertise on the part of the reader in order to differentiate Pneumocystis jiroveci from staining artifacts and other fungi. This real-time PCR assay provides sensitive (21% more sensitive than direct detection using fluorescent calcofluor white stain), specific, and objective detection of Pneumocystis from bronchoalveolar lavage fluid and other specimens.

Reference Values Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Not applicable

Interpretation Provides information to assist in interpretation of the test results

A positive result indicates the presence of Pneumocystis DNA.

 

A negative result indicates the absence of detectable Pneumocystis DNA.

Cautions Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Test results should be used as an aid in diagnosis and should not be considered diagnostic in themselves. The literature indicates that Pneumocystis can cause asymptomatic colonization of healthy and immunocompromised individuals. Therefore, test results should be correlated with patient symptoms and clinical presentation.

 

A negative result does not rule out the presence of Pneumocystis or active disease because the organism may be present at undetectable levels.

Supportive Data

A total of 221 bronchoalveolar lavage (BAL) fluid samples were evaluated for the presence of Pneumocystis DNA by the LightCycler and compared to fluorescent microscopy using calcofluor white staining. Of the 221, 24 were positive and 190 were negative by both detection methods. The remaining 7 were positive by PCR and negative by microscopy. The 7 specimens that were positive using LightCycler PCR alone were tested using another PCR assay targeting a second Pneumocystis gene. All 7 specimens were positive using the second target suggesting that they were true positives that were undetected using the microscopic method. The sensitivity, specificity, positive and negative predictive values of this real-time PCR assay is 100%, 96%, 77%, and 100%, respectively. The analytical sensitivity of the method is 5.6 copies/ul of positive plasmid control or approximately 28 copies/reaction. The analytical sensitivity in spiked, pooled BAL specimens was found to be 56 targets/uL using the positive control plasmid. PCR inhibition was tested by spiking 50 extracted negative respiratory specimens (including 10 BAL fluid specimens) with 100 copies of target/ul using a positive control plasmid. No PCR inhibition was detected. The specificity of the PCR assay was determined by evaluating DNA extracted from pure cultures of a variety of bacteria and fungi. Extracted human DNA was analyzed as well. None of the microbial or human DNA was amplified by the Pneumocystis LightCycler assay indicating that the assay is specific for Pneumocystis species.

Clinical Reference Provides recommendations for further in-depth reading of a clinical nature

1. Cushion MT: Pneumocystis. In Manual of Clinical Microbiology. 8th edition. Edited by PR Murray, EJ Baron, JH Jorgensen, et al: Washington, DC, ASM Press, 2003, pp 1712-1725

2. Maskell NA, Waine DJ, Lindley A, et al: Asymptomatic carriage of Pneumocystis jiroveci in subjects undergoing bronchoscopy: a prospective study. Thorax July 2003;58(7):594-597

3. Miller RF, Ambrose HE, Wakefield AE: Pneumocystis carinii f. sp. hominis DNA in immunocompetent health care workers in contact with patients with P. carinii pneumonia. J. Clin. Microbiol November 2001;39(11):3877-3882

4. Takahashi T, Goto M, Endo T, et al: Pneumocystis carinii carriage in immunocompromised patients with and without human immunodeficiency virus infection. J Med Microbiol July 2002;51(7):611-614

5. Vargas SL, Hughes WT, Santolaya ME, et al: Search for primary infection by Pneumocystis carinii in a cohort of normal, healthy infants. Clin Infect Dis March 2001;32(6):855-861

6. Wakefield AE, Lindley AR, Ambrose HE, et al: Limited asymptomatic carriage of Pneumocystis jiroveci in human immunodeficiency virus-infected patients. J Infect Dis 2003;187(6):901-908

Method Description Describes how the test is performed and provides a method-specific reference

Bronchoalveolar lavage fluid and sputum are liquified using N-acetyl-L-cysteine. Following liquification, cells are pelleted by centrifugation and resuspended in S.T.A.R. buffer (Roche). Nucleic acids are extracted using the MagNA Pure LC Instrument (Roche). The extract is then transferred to individual self-contained cuvettes for amplification using the LightCycler real-time PCR platform (Roche). The LightCycler is an automated instrument that amplifies and monitors the development of target nucleic acid (amplicon) after each cycle of PCR. The detection of amplicon is based on fluorescence resonance energy transfer (FRET), which utilizes hybridization probes. The presence of the specific organism nucleic acid is confirmed by performing a melting curve analysis of the amplicon. (Arcenas RC, Uhl JR, Buckwalter SP, et al., A Real-Time PCR Assay for Detection of Pneumocystis from Bronchoalveolar Lavage Fluid. Diagn Microbiol Infect Dis, 2006, 54:169-175)

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday through Friday

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

Same day/1 day

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result

1 day

Specimen Retention Time Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

3 to 14 days

Performing Laboratory Location The location of the laboratory that performs the test

Rochester

Test Classification Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the U.S. Food and Drug Administration.

CPT Code Information Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

87798

LOINC® Code Information Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.

Result IDReporting NameLOINC Code
SRC63Specimen Source31208-2
81698Result49441-9
24188Special Information48767-8
24189Report StatusN/A