Mobile Site ›
Normal View

Test ID: IOD
Iodine, Serum

Secondary ID A test code used for billing and in test definitions created prior to November 2011

81574

NY State Approved Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Useful For Suggests clinical disorders or settings where the test may be helpful

Determination of iodine overload

 

Monitoring iodine levels in individuals taking iodine-containing drugs

Special Instructions and Forms Describes specimen collection and preparation information, test algorithms, and other information pertinent to test. Also includes pertinent information and consent forms to be used when requesting a particular test

Method Name A short description of the method used to perform the test

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)

Reporting Name A shorter/abbreviated version of the Published Name for a test; an abbreviated test name

Iodine, S

Specimen Type Describes the specimen type needed for testing

Serum

Specimen Required Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.

Container/Tube: Plain, royal blue-top Monoject trace element blood collection tube, product #8881-307006 (Supply T184)

Specimen Volume: 1 mL

Collection Instructions: See Metals Analysis-Collection and Transport in Special Instructions for complete instructions.

Additional Information: High concentrations of gadolinium are known to interfere with most metals tests. If gadolinium-containing contrast media has been administered, a specimen should not be collected for 48 hours.

Forms: If not ordering electronically, submit a General Request Form (Supply T239) with the specimen.

Specimen Minimum Volume Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.

0.3 mL

Reject Due To Identifies specimen types and conditions that may cause the specimen to be rejected

Hemolysis

Mild OK; Gross OK

Lipemia

Mild OK; Gross OK

Icterus

Mild OK; Gross reject

Other

NA

Specimen Stability Information Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.

Specimen TypeTemperatureTime
SerumRefrigerated (preferred)21 days
 Ambient 21 days
 Frozen 21 days

Clinical Information Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Iodine is an essential element that is required for thyroid hormone production. The measurement of iodine serves as an index of adequate dietary iodine intake and iodine overload, particularly from iodine-containing drugs such as amiodarone.

Reference Values Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

40-92 ng/mL

Interpretation Provides information to assist in interpretation of the test results

Values between 80 ng/mL and 250 ng/mL have been reported to indicate hyperthyroidism.

 

Values >250 ng/mL may indicate iodine overload.

Cautions Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

There are no known analytical interferences with this procedure.

 

Administration of iodine-containing contrast media will yield elevated results.

 

Disinfectants (such as Betadine) that contain iodine should not be used during venipuncture.

 

High concentrations of gadolinium are known to interfere with most metals tests. If gadolinium-containing contrast media has been administered, a specimen should not be collected for 48 hours.

Clinical Reference Provides recommendations for further in-depth reading of a clinical nature

Allain P, Berre S, Krari N, et al: Use of plasma iodine assay for diagnosing thyroid disorders. J Clin Pathol 1993 May;46(5):453-455

Method Description Describes how the test is performed and provides a method-specific reference

This assay is performed on an inductively coupled plasma-mass spectrometer. Calibrating standards and blanks are diluted with an aqueous basic diluent containing internal standards. Quality control specimens and patient samples are diluted in an identical manner. In turn, all diluted blanks, calibrating standards, quality control specimens, and patient specimens are aspirated into a pneumatic nebulizer and the resulting aerosol directed to the hot plasma discharge by a flow of argon. In the annular plasma, the aerosol is vaporized, atomized, then ionized. The ionized gases plus neutral species formed in the annular plasma space are aspirated from the plasma through an orifice into a quadrupole-mass spectrometer. The mass range from 1 amu to 263 amu is rapidly scanned multiple times and ion counts tabulated for each mass of interest. Instrument response is defined by the linear relationship of analyte concentration versus ion count ratio (analyte ion count/internal standard ion count). Analyte concentrations are derived by reading the ion count ratio for each mass of interest and determining the concentration from the response line.(Unpublished Mayo method)

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday, Tuesday, Wednesday, Friday; 5 p.m.

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

1 day

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result

3 days

Specimen Retention Time Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

2 weeks

Performing Laboratory Location The location of the laboratory that performs the test

Rochester

Test Classification Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the U.S. Food and Drug Administration.

CPT Code Information Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

83789

LOINC® Code Information Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.

Result IDReporting NameLOINC Code
81574Iodine, S2494-3