Test ID: MHRP
Mycoplasma hominis, Molecular Detection, PCR

NY State Approved Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Useful For Suggests clinical disorders or settings where the test may be helpful

Rapid, sensitive, and specific identification of Mycoplasma hominis from synovial fluid, genitourinary, reproductive, and lower respiratory sources

Method Name A short description of the method used to perform the test

Real-Time Polymerase Chain Reaction (PCR) Using LightCycler and Fluorescent Resonance Energy Transfer (FRET)
(PCR is utilized pursuant to a license agreement with Roche Molecular Systems, Inc.)

Reporting Name A shorter/abbreviated version of the Published Name for a test; an abbreviated test name

Mycoplasma hominis PCR

Specimen Type Describes the specimen type needed for testing

Varies

Specimen Required Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.

Forms: If not ordering electronically, submit a Microbiology Request Form (Supply T244) with the specimen.

 

Specimen source is required.

 

Submit only 1 of the following specimens:

 

Specimen Type: Swab

Sources: Cervix, urethra, urogenital, vaginal

Container/Tube:                              

Preferred: Dacron or rayon swab with aluminum or plastic shaft (does not need to be in transport media)

Acceptable: Swab in transport media: M5 (Supply T484), M4, or universal transport media

Specimen Volume: Entire specimen    

Collection Instructions:

Vaginal: Collect specimen by swabbing back and forth over mucosa surface to maximize recovery of cells.

Urethra or Cervical: Collect specimen by inserting swab 1 to 3 cm and rotating 360 degrees.

 

Specimen Type: Fluid

Sources: Pelvic, amniotic, prostatic secretion, semen, reproductive drainage, lower respiratory, synovial

Container/Tube:                                 

Preferred: Sterile container

Acceptable: Container with 3 mL of transport media: M5 (Supply T484), M4, or universal transport media

Specimen Volume: 1-2 mL

 

Specimen Type: Urine (kidney stones)

Container/Tube: Sterile container

Specimen Volume: 10 mL

 

Specimen Type: Tissue

Sources: Placenta, products of conception, genitourinary

Container/Tube: Sterile container

Specimen Volume: 5 mm

Collection Instructions: Submit only fresh tissue.

Specimen Minimum Volume Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.

Pelvic Fluid, Amniotic Fluid, Prostatic Secretions, Semen, Reproductive Drainage or Fluid, Lower Respiratory, Synovial Fluid: 1 mL/Urine: 2 mL/Swab: 1 swab

Reject Due To Identifies specimen types and conditions that may cause the specimen to be rejected

Hemolysis

NA

Lipemia

NA

Icterus

NA

Other

Wooden shaft, cotton swab

Specimen Stability Information Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.

Specimen TypeTemperatureTime
VariesRefrigerated (preferred)7 days
 Frozen 7 days

Clinical Information Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Mycoplasma hominis causes septic arthritis and postpartum fever, and has been associated with pelvic inflammatory disease and bacterial vaginosis. Culture isolation is laborious, requiring a high degree of technical skill and taking several days. PCR detection is sensitive, specific, and provides same-day results. The described PCR assay has replaced conventional culture for Mycoplasma hominis at Mayo Medical Laboratories.

Reference Values Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Not applicable

Interpretation Provides information to assist in interpretation of the test results

A positive PCR result for the presence of a specific sequence found within the Mycoplasma hominis tuf gene indicates the presence of Mycoplasma hominis DNA in the specimen.

 

A negative PCR result indicates the absence of detectable Mycoplasma hominis DNA in the specimen, but does not rule-out infection as false-negative results may occur due to the following; inhibition of PCR, sequence variability underlying the primers and/or probes, or the presence of Mycoplasma hominis in quantities less than the limit of detection of the assay.

Cautions Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Interfering substances may affect the accuracy of this assay; results should always be interpreted in conjunction with clinical and epidemiological findings.

 

This test does not detect other mycoplasmas or ureaplasmas (including Mycoplasma pneumoniae, a common cause of community acquired pneumonia).

Supportive Data

This assay was clinically validated in a blinded manner using 292 archived specimens submitted for Mycoplasma hominis culture, 281 of which were in M4, M5, or Universal Transport Media (UTM). The specimens consisted of 251 genitourinary, 32 reproductive fluids or tissues, and 9 respiratory specimens. Compared to culture, PCR had 90.7% (39/43) sensitivity and 99.2% (245/247) specificity (p=0.41). Discordant results on the archived specimens (n=6) were tested by PCR by Dr. Kathleen A. Stellrecht at the Albany Medical Center (Journal of Clinical Microbiology 2004;42:1528-1533). Dr. Stellrecht found both of the PCR-positive/culture-negative specimens to be PCR positive, and 3 of 4 specimens that were PCR negative/culture positive to be PCR negative. The limit of detection of the assay is 100 targets/mcl for all validated sources.  Additional whole organism spiking studies (at the limit of detection of the assay) were performed using 30 or more of each of the following: genitourinary swabs, respiratory specimens, genitourinary fluid, reproductive tissue/fluid and synovial fluid. All results (except for 6) were as expected.

Clinical Reference Provides recommendations for further in-depth reading of a clinical nature

1. Waites KB, Taylor-Robinson D: Mycoplasma and Ureaplasma. In Manual of Clinical Microbiology. 10th edition. Edited by J Versalovic. ASM Press, Washington, DC, 2011

2. Stellrecht KA, Woron AM, Mishrik NG, Venezia RA: Comparison of multiplex PCR assay with culture detection of genital mycoplasmas. J Clin Microbiol 2003;42:1528-1533

3. Kenny GE: Genital mycoplasmas: Mycoplasma genitalium, Mycoplasma hominis, and Ureaplasma species. In Mandell, Douglas, and Bennett's Principles and Practice of Infectious Diseases. Edited by GL Mandell, et al. Churchill Livingstone, New York, 2008

4. Yoshida T, Maeda SI, Deguchi T, et al: Rapid detection of Mycoplasma genitalium, Mycoplasma hominis, Ureaplasma parvum, and Ureaplasma urealyticum organisms in genitourinary samples by PCR-microtiter plate hybridization assay. J Clin Microbiol 2003;41:1850-1855

5. Cunningham SA, Shannon SK, Rosenblatt J, Patel R: Comparison of culture and rapid polymerase chain reaction detection of Mycoplasma hominis, Ureaplasma urealyticum and Ureaplasma parvum from genitourinary specimens. 48th Annual Infectious Diseases Society of America Meeting, 2010

Method Description Describes how the test is performed and provides a method-specific reference

This PCR method employs a target-specific detection system including primers, as well as fluorescent resonance energy transfer (FRET) hybridization probes designed for tuf of Mycoplasma hominis. The LightCycler instrument amplifies and monitors target nucleic acid sequences by fluorescence during PCR cycling. This is an automated PCR system that can rapidly detect amplified product development. The detection of amplified products is based on the FRET principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3' end is excited by an external light source, which emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, on the 5' end. The acceptor fluorophore then emits light of a different wavelength that is measured with a signal that is proportional to the amount of specific PCR product. The process is completed in a closed-tube system.(Unpublished Mayo method)

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday through Friday

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

3 days

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result

4 days

Specimen Retention Time Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location The location of the laboratory that performs the test

Rochester

Test Classification Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the U.S. Food and Drug Administration.

CPT Code Information Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

87798

LOINC® Code Information Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.

Result IDReporting NameLOINC Code
SRC86Specimen source31208-2
32536ResultN/A