Test ID: HGRU
Mercury, Random, Urine
Secondary ID
A test code used for billing and in test definitions created prior to November 2011
NY State Approved
Indicates the status of NY State approval and if the test is orderable for NY State clients.
Useful For
Suggests clinical disorders or settings where the test may be helpful
Detecting mercury toxicity
Special Instructions and Forms
Describes specimen collection and preparation information, test algorithms, and other information pertinent to test. Also includes pertinent information and consent forms to be used when requesting a particular test
Method Name
A short description of the method used to perform the test
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
Reporting Name
A shorter/abbreviated version of the Published Name for a test; an abbreviated test name
Aliases
Lists additional common names for a test, as an aid in searching
Mercury (Hg)
Specimen Type
Describes the specimen type needed for testing
Specimen Required
Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.
Collection Container/Tube: Clean, plastic urine collection container
Submission Container/Tube: Plastic, 6-mL tube (Supply T465) or a clean, plastic aliquot container with no metal cap or glued insert
Specimen Volume: 2 mL
Collection Instructions:
1. Collect a random urine specimen.
2. See Metals Analysis-Collection and Transport in Special Instructions for complete instructions.
Additional Information: High concentrations of gadolinium and iodine are known to interfere with most metals tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Specimen Minimum Volume
Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.
Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected
| Hemolysis | NA |
| Lipemia | NA |
| Icterus | NA |
| Other | NA |
Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.
| Specimen Type | Temperature | Time |
|---|---|---|
| Urine | Refrigerated (preferred) | 7 days |
| Frozen | 7 days |
Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test
The correlation between the levels of excretion in the urine and the clinical symptoms is considered poor, but urinary mercury (Hg) is the most reliable way to assess exposure to inorganic mercury.
For additional information, see HG/8618 Mercury, Blood.
Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.
0-15 years: not established
> or =16 years: 0-9 mcg/L
Interpretation
Provides information to assist in interpretation of the test results
Daily urine excretion >50 mcg/day indicates significant exposure (per World Health Organization standard).
Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances
To avoid contamination by dust, the specimen should be collected away from the site of suspected exposure.
High concentrations of gadolinium and iodine are known to interfere with most metals tests. If gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Clinical Reference
Provides recommendations for further in-depth reading of a clinical nature
1. Lee R, Middelton D, Caldwell K, et Al: A Review of Events That Expose Children to Elemental Mercury in the United States. Environ Health Perspect 2009;117:871-878
2. Bjorkman L, Lundekvam BF, Laegreid T, et al: Mercury in human brain, blood, muscle and toenails in relation to exposure: an autopsy study. Environmental Health 2007;6:1186-1476
Method Description
Describes how the test is performed and provides a method-specific reference
This assay is performed on an inductively coupled plasma-mass spectrometer. Calibrating standards and blanks are diluted with an aqueous acidic diluent containing internal standard(s). Quality control specimens and patient samples are diluted in an identical manner. In turn, all diluted blanks, calibrating standards, quality control specimens, and patient specimens are aspirated into a pneumatic nebulizer and the resulting aerosol directed to the hot plasma discharge by a flow of argon. In the annular plasma the aerosol is vaporized, atomized, then ionized. The ionized gases plus neutral species formed in the annular plasma space are aspirated from the plasma through an orifice into a quadrupole mass spectrometer. The mass range from 1 amu to 263 amu is rapidly scanned multiple times and ion counts tabulated for each mass of interest. Instrument response is defined by the linear relationship of analyte concentration vs. ion count ratio (analyte ion count/internal standard ion count). Analyte concentrations are derived by reading the ion count ratio for each mass of interest and determining the concentration from the response line. (Nixon DE, Burritt MF, Moyer TP: The determination of mercury in whole blood and urine by inductively coupled plasma mass spectrometry. Spectrochimica Acta, Part B-Atomic Spectroscopy 1999;54:1141-1153 and Hanley MM, Eckdahl SJ, Kiedrowski B, Moyer TP, Murray DL. A comparison of methods for attenuation of oxide interferences in cadmium and mercury analysis by ICP-MS. (Abstract 627) 38th Federation of Analytical Chemistry and Spectroscopy Societies, Reno, Nevada, October 2-6, 2011)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Monday through Friday; 3 p.m. - 9 p.m.; Saturday; 8 a.m. – 3 p.m./Continuously
Analytic Time
Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.
Maximum Laboratory Time
Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result
Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded
Performing Laboratory Location
The location of the laboratory that performs the test
Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.
CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.
83825
LOINC® Code Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.
| Result ID | Reporting Name | LOINC Code |
|---|---|---|
| 60248 | Mercury, Random, U | 5689-5 |