Mobile Site ›
Normal View

Test ID: PBRU
Lead, Random, Urine

Secondary ID A test code used for billing and in test definitions created prior to November 2011

60246

NY State Approved Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Useful For Suggests clinical disorders or settings where the test may be helpful

Detecting clinically significant lead exposure

Special Instructions and Forms Describes specimen collection and preparation information, test algorithms, and other information pertinent to test. Also includes pertinent information and consent forms to be used when requesting a particular test

Method Name A short description of the method used to perform the test

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)

Reporting Name A shorter/abbreviated version of the Published Name for a test; an abbreviated test name

Lead, Random, U

Aliases Lists additional common names for a test, as an aid in searching

Lead (Pb)
Pb (Lead)

Specimen Type Describes the specimen type needed for testing

Urine

Specimen Required Defines the optimal specimen. This field describes the type of specimen required to perform the test and the preferred volume to complete testing. The volume allows automated processing, fastest throughput and, when indicated, repeat or reflex testing.

Collection Container/Tube: Clean, plastic urine collection container

Submission Container/Tube: Plastic, 6-mL tube (Supply T465) or a clean, plastic aliquot container with no metal cap or glued insert

Specimen Volume: 2 mL

Collection Instructions:

1. Collect a random urine specimen.

2. See Metals Analysis-Collection and Transport in Special Instructions for complete instructions.

Additional Information: High concentrations of gadolinium and iodine are known to interfere with most metals tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.

Specimen Minimum Volume Defines the amount of specimen required to perform an assay once, including instrument and container dead space. Submitting the minimum specimen volume makes it impossible to repeat the test or perform confirmatory or perform reflex testing. In some situations, a minimum specimen volume may result in a QNS (quantity not sufficient) result, requiring a second specimen to be collected.

2 mL

Reject Due To Identifies specimen types and conditions that may cause the specimen to be rejected

Hemolysis

NA

Lipemia

NA

Icterus

NA

Other

NA

Specimen Stability Information Provides a description of the temperatures required to transport a specimen to the laboratory. Alternate acceptable temperature(s) are also included.

Specimen TypeTemperatureTime
UrineRefrigerated (preferred)28 days
 Ambient 28 days
 Frozen 28 days

Clinical Information Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Increased urine lead concentration indicates significant lead exposure.

 

Measurement of urine lead concentration before AND after chelation therapy has been used as an indicator of significant lead exposure. An increase in lead concentration in the postchelation specimen of up to 6 times the concentration in the prechelation specimen is normal. Blood lead is the best clinical correlate of toxicity.

 

For additional information, see PBBD/15070 Lead with Demographics, Blood.

Reference Values Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

0-4 mcg/L

Reference values apply to all ages.

Interpretation Provides information to assist in interpretation of the test results

Urinary excretion of <4 mcg/L is not associated with any significant lead exposure.

 

Urinary excretion >4 mcg /L is usually associated with pallor, anemia, and other evidence of lead toxicity.

Cautions Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

This test is not a substitute for blood lead screening.

 

High concentrations of gadolinium and iodine are known to interfere with most metals tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.

Clinical Reference Provides recommendations for further in-depth reading of a clinical nature

1. Kosnett MJ, Wedeen RP, Rotherberg SJ, et al: Recommendations for medical management of adult lead exposure. Environ Health Perspect 2007;115:463-471

2. de Burbane C, Buchet JP, Leroyer A, et al: Renal and neurologic effects of cadmium, lead, mercury, and arsenic in children: evidence of early effects and multiple interactions at environmental exposure levels.Environ Health Perspect 2006;114:584-590

3. Pascal DC, Ting BG, Morrow JC, et al: Trace metals in urine of United States residents: reference range concentrations.Environ Res 1998;76(1):53-59

Method Description Describes how the test is performed and provides a method-specific reference

This assay is performed on an inductively coupled plasma-mass spectrometer. Calibrating standards and blanks are diluted with an aqueous acidic diluent containing internal standard(s). Quality control specimens and patient samples are diluted in an identical manner. In turn, all diluted blanks, calibrating standards, quality control specimens and patient specimens are aspirated into a pneumatic nebulizer and the resulting aerosol directed to the hot plasma discharge by a flow of argon. In the annular plasma the aerosol is vaporized, atomized, then ionized. The ionized gases plus neutral species formed in the annular plasma space are aspirated from the plasma through an orifice into a quadrupole mass spectrometer. The mass range from 1 amu to 263 amu is rapidly scanned multiple times and ion counts tabulated for each mass of interest. Instrument response is defined by the linear relationship of analyte concentration vs. ion count ratio (analyte ion count/internal standard ion count) Analyte concentrations are derived by reading the ion count ratio for each mass of interest and determining the concentration from the response line. (Nixon DE, Moyer TP: Routine clinical determination of lead, arsenic, cadmium, and thallium in urine and whole blood by inductively coupled plasma mass spectrometry. Spectrochimica Acta Part B-Atomic Spectroscopy 1996;1:13-25)

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday through Friday; 3 p.m. – 9 p.m.; Saturday; 8 a.m. – 3 p.m./Continuously

 

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

1 day

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result

2 days

Specimen Retention Time Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

14 days

Performing Laboratory Location The location of the laboratory that performs the test

Rochester

Test Classification Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer's instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR), Investigation Use Only (IUO) product, or a Research Use Only (RUO) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the U.S. Food and Drug Administration.

CPT Code Information Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Medical Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

83655

LOINC® Code Information Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the result codes returned for this test or profile.

Result IDReporting NameLOINC Code
60246Lead, Random, U5676-2