Test ID: INHB
Inhibin B, Serum
Method Description
Describes how the test is performed and provides a method-specific reference
Inhibin B Gen II ELISA (enzyme-linked immunosorbent assay) is an enzymatically amplified 3-step "sandwich'' assay. Sample is incubated in wells that have been coated with antiactivin B antibody. After incubation and washing, the wells are incubated with biotinylated anti-inhibin detection antibody. After a second incubation and washing step, the wells are incubated with streptavidin labeled with the enzyme horseradish peroxidase. Antibody-analyte complex is detected by dual wavelength absorbance measurement after addition of the tetramethylbenzidine substrate. The absorbance measured is directly proportional to the concentration of inhibin B in the samples. (Package insert: Inhibin B Gen II ELISA kit, Beckman Coulter, Inc. Brea, CA)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Tuesday, Thursday; 9:00 a.m.