Test ID: AHEPR    
Acute Hepatitis Profile with Hepatitis C Virus Reflex

Method Description Describes how the test is performed and provides a method-specific reference

Hepatitis B Surface Antigen (HBsAg):

Specimens are first tested by the VITROS HBsAg assay. With modification to the assay manufacturer's instructions for use, specimens yielding S/CO > or =1.00 but < or =50.0 will be confirmed by the VITROS HBsAg Confirmatory assay. Specimens that are strongly positive (ie, S/CO >50.0) do not require this confirmation.

 

This immunometric technique involves the simultaneous reaction of HBsAg in the sample with mouse monoclonal anti-HBs antibody coated onto the wells and a horseradish peroxidase (HRP)-labeled mouse monoclonal anti-HBs antibody in the conjugate. Unbound conjugate is removed by washing. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent increases the level and duration of the light produced. The light signals are read by the VITROS ECi System. The amount of HRP conjugate bound is indicative of the level of HBsAg present in the sample.(Package insert: VITROS HBsAg assay, no. J03798, version 1.0; Ortho-Clinical Diagnostics, Inc. Rochester, NY)

 

HBsAg Confirmation:

The VITROS HBsAg Confirmatory Kit uses the principle of specific antibody neutralization to confirm the presence of HBsAg. The sample is tested twice: 1 aliquot is incubated with a neutralizing reagent containing high titer anti-HBs (the confirmatory antibody); the second aliquot is incubated with a non-neutralizing control reagent (the sample diluent). The confirmatory antibody binds to HBsAg in the sample inhibiting its reaction in the VITROS HBsAg assay. This leads to a reduced result compared to that for the non-neutralized control sample.(Package insert: VITROS HBsAg Confirmation assay, no. J10583, version 1.0; Ortho-Clinical Diagnostics, Inc. Rochester, NY)

 

Hepatitis A IgM Antibody:

An antibody class capture technique is used. This involves the dilution of the sample and the simultaneous reaction of IgM in the diluted sample with biotinylated mouse monoclonal antihuman IgM antibody. The immune complex is captured by streptavidin on the wells. Unbound materials are removed by washing. Horseradish peroxidase (HRP)-labeled mouse monoclonal anti-hepatitis A virus (anti-HAV) IgM, which has been complexed with recombinant HAV antigen (conjugate), is then captured by anti-HAV-specific IgM bound to the wells. Unbound material is removed by washing.

 

The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent, is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent increases the level and duration of the light produced. The light signals are read by the VITROS ECi system. The amount of HRP conjugate is indicative of the concentration of anti-HAV IgM present in the sample.(Package insert: VITROS Anti-HAV IgM Reagent Pack, no. GEM1235A, version 3.0; Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626-5101, 8/18/2009)

 

Hepatitis B Core (HBc) IgM Antibody:

An antibody class capture technique is used. This involves the dilution of the sample and the simultaneous reaction of IgM in the diluted sample with biotinylated mouse monoclonal antihuman IgM antibody. The immune complex is captured by streptavidin on the wells. Unbound materials are removed by washing. Horseradish peroxidase (HRP)-labeled mouse monoclonal anti-hepatitis B core (anti-HBc) IgM antibody, which has been complexed with recombinant HBc antigen (conjugate), is then captured by anti-HBc specific IgM bound to the wells. Unbound material is removed by washing.

 

The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminal derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminal derivative, producing light. The electron transfer agent increases the level and duration of the light produced. The light signals are read by the VITROS ECi System. The amount of HRP conjugate bound is indicative of the concentration of anti-HBc IgM present in the sample.(Package insert: VITROS Anti-HBc IgM assay, Ortho-Clinical Diagnostics, Raritan, NJ)

 

Hepatitis C Virus (HCB) Antibody Screen:

The VITROS anti-HCV assay is performed using the VITROS Anti-HCV Reagent Pack and VITROS Immunodiagnostic Products Anti-HCV Calibrator on the VITROS 3600 Immunodiagnostic System (Ortho-Clinical Diagnostics, Inc., Raritan, NJ). An immunometric technique is used, involving a 2-stage reaction. In the first stage, HCV antibody present in the sample binds to HCV recombinant antigens coated on the reaction wells, and unbound sample is removed by washing. In the second stage, horseradish peroxidase (HRP)-labeled antibody conjugate (mouse monoclonal antihuman IgG) binds to human IgG captured on the well in the first stage. Unbound conjugate is removed by washing. A reagent containing luminogenic substrates (a luminal derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminal derivative, producing light. The electron transfer agent increases the level and duration of the light produced. The emitted light signals are detected and measured by the VITROS 3600 Immunodiagnostic System. The amount of HRP conjugate bound is directly proportional to the level of anti-HCV antibodies present in a given sample.(Ismail N, Fish GE, Smith MN: Laboratory evaluation of a fully automated chemiluminescence immunoassay for rapid detection of HBsAg, antibodies to HBsAg, and antibodies to hepatitis C virus. J Clin Microbiol 2004;42:610-617)

 

See individual test ID for reflex tests.

Supplemental Report Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday through Saturday; varies

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

1 day

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Medical Laboratories until the release of the test result

2 days

Specimen Retention Time Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location The location of the laboratory that performs the test

Rochester