150 uL of each serum sample is diluted with 150 uL of reagent-grade
water and 2,700 uL of acidic diluent (1% HNO[3]) containing gallium,
cobalt, and yttrium internal standards. The diluted samples are then
vortexed and analyzed via dynamic reaction cell-inductively coupled
plasma-mass spectrometer (DRC-ICP-MS). The sample's raw
intensity is compared to a plasma matrix-matched calibration curve,
which spans from 0 ng/mL Mn spike to 10.90 ng/mL Mn spike
standards. Gallium, cobalt, and yttrium are combined as the internal
standard for manganese analysis, and the intensity of manganese
is adjusted according to the intensities of the internal standard.
(Unpublished Mayo method)