Test ID: FRDIG
Digoxin, Free, Serum
Method Description
Describes how the test is performed and provides a method-specific reference
Free digoxin not bound to the digoxin-specific antibody fragments and not bound to protein is separated from bound digoxin by a 30-kD centrifugal filter device. Free digoxin passes through the filter; bound digoxin is retained in the filter. The filtrate is analyzed for digoxin using the E170 Digoxin method, a competitive electrochemiluminescence immunoassay that employs a monoclonal antibody directed against digoxin. Digoxin in the specimen competes with the added digoxin derivative labeled with biotin for the binding sites on the ruthenylated antibody-complex. Streptavidin-coated microparticles are added and the mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Application of voltage to the electrode induces the chemiluminescent emission, which is then measured.(Package insert: Roche Digoxin, 2005)
Day(s) and Time(s) Test Performed
Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.
Monday through Sunday; Continuously