Unit Code 81369:
Fatty Acid Profile, Peroxisomal (C22-C26), Serum
Method Description
Acidic hydrolysis is followed by basic hydrolysis and reacidification.
Hexane extraction then proceeds to derivatization with
pentafluorobenzyl bromide (PFB). Separation and detection of PFB-esters
is accomplished by capillary gaschromatograph-mass spectrometry
(GC/MS) using electron capture ionization and selected negative ion
monitoring. Quantitation is enhanced by the use of stable isotope-labeled
internal standards. (Stellard F, ten Brink HJ, Kok RM, et al: Stable isotope
dilution analysis of very long chain fatty acids in plasma, urine and
amniotic fluid by electron capture negative ion mass fragmentography.
Clin Chim Acta1990;192:133-144)
Performing Laboratory Location
Rochester