Haloperidol, reduced haloperidol (its major metabolite), and the internal
standard (haloperidol d4) are extracted from alkaline serum into 5%
isopropyl alcohol in hexane. The organic fraction is dried and reconstituted
with methanol. Analysis is performed on an isocratic liquid chromatography
(LC) system with tandem mass spectrometry (MS/MS) detection. (Hempenius J,
Steenvoorden RJ, Lagerwerf FM, et al: "High throughput†solid-phase
extraction technology and turbo ionspray LC-MS-MS applied to the deter-
mination of haloperidol in human plasma. J Pharmaceut Biomed Anal
1999;20:889-898)